9 Repair Procedures
9.22 Control module
To replace the fuses
and 由e
fan
fll幅『
Toremove 曲eωntrol
module
Ml110140
08/07
Th e control module is mounted on either a folding mount oron a display arm. In addition
to the power
co时,
signals between the control module and the machine mounted
components are sent through three cables that pass through the side of the machine
and connectto the rear ofthe control module.
To service the color display , refer to Section 9.22.1.
To service the original display , referto Section 9.22.2.
<
CM)
To replace the fuses (A) , first remove the power cord retainer (8)
and 由epower
cord (C).
To replace the filter (D) , remove the filter retainer
(罔.Ifthe
Control Module is
mounted on an arm , remove the control panel from the arm to access the filter.
1. Disconnectthe three interface cables
(町 and
the power
co时 (C).
2. Remove the screws that attach the control module
to 由e
mounting brackets.
3. Remove the control module (CM).
4. To replace the control module , reassemble in reverse order.
5.
ReferωSection
8.3 for
ins阳 lling
software and updating the control board
configuration.
9-55