Error Log - GE Aespire 7100 Technical Reference Manual

Anestesia machine
Table of Contents

Advertisement

4.5 Error Log
4 Tests and Calibration
Th ere are two special types of alarms:
Minimum monitoringalarms thatstop mechanical ventilation
Minimum shutdown alarms that stop mechanical ventilation and monitoring.
An alarm message that results from these special types of alarms is considered an error
alarm.
Th e Errol Log displays up to 20 ofthe most current Error messages that have been
logged.
Each log entry shows:
Bootup Count number (the bootup count is incremented each time the
machine is turned on).
T1 me (ms) - the time in milliseconds since bootup when the latest alarm
condition occurred.
Erro
r message associated with the particular alarm condition.
Address - the place in the software sequence where the last occurrence ofthe
errorωokplace.
Th e error address and software revision are important pieces of information to note if
technical
suppo此 is
required.
Th e
bo忱。m-Ieft
comer of the screen displays additional information:
Bootup Count
La
st Cleared
Current Bootup Count
口幅rEπOIL吨
自lOIUP C四ntL甜 CI剧创 1
0町田11 Boot叩 E国mt:
55
Go to SelV lce Modes
M酬』
Sc roll
Data
To view all the log entries
,
select "Scroll Data."
Th en , use the control knob
ωscroll
through the entries.
To exit Scroll Data , press the control knob.
Clear Error Log
To clearthe error log
,
select "Clear Error
Log." 而esystem as陆 youωconfirm
thatyou
want to clear the error log.
Ml110140 08/07
Remarks
Afteryou clearthe alarm log:
the "Boot Count
La
st Cleared" numberwill be resetto the "Current Boot Count"
number.
the menu will showthe message "No entries in log!".
4-7