9. Install the replacementswitch assembly:
a.
Lo
osen the two outside screws on the electrical module.
b. Insert the wires in the electrical module and tighten the screws.
c. Pull the wires on the electrical module
to
ensure thatthere is a good connection.
d. Turn back the system switch mounting screws until their tips recede.
e. Orientthe switch assembly with the Air
fi忧ings towarcl 由 e
right and the O
2
fittings towarcl the left.
f. Install the switch assembly through the gauge pane l.
g.
Push 由 e
knob collar in
wi由 the
indicator up and turn it clockwise
un刨 itloc阳.
h. Tighten the mounting screws. Make sure that the top edge of the switch
assembly is parallel to the top edge ofthe gauge pane l.
i. Transfer the tubing from the old system switch to the new system switch on the
pneumatic module (pull on the tubing to
ensure 伽 at
it is locked
into 由e
module).
02 0ut (PO此 4)
A:::::::ED皿L
02 1n (P由此 3)
(Rear Vl棚}
10. Testthe replacementswitch assembly:
a. ConnectAir and O
2
supplies.
b. Connect the power cable to an electrical outlet
c. Set the system switch to On.
d. Increasethe O
2
and Airflow. Makesure thatgas flows.
e. Make sure that you do not feel or hear any leaks.
f. Make sure that the display comes On.
g. Setthe system switch to Standby.
h. Make sure all gas flow stops and the display tums
0侃
11. Reinstall the gauge panel and the
tableωp.
12. Perform the checkout procedure (Section 3).