Testsetup; Testin G/Ad Justing Secondary Regulators Or Balance Regulators - GE Aespire 7100 Technical Reference Manual

Anestesia machine
Table of Contents

Advertisement

Aespire 7100
5.1 Primary
Regulato附
5- 2
Follow the
procedu 陀 in
Section 5. 1.1 to gain access to the regulators. Th en
,
in Section 5. 1. 2 , selectthe testthat is appropriate for the regulator you are
testing.
~
WARNING
When testingjadjusting N
2
0 regulators
,
nitrous oxide flows through the
system. Use a safe and approved procedure to collect and remove i
t.
5.1.1 Test setup
&.
WARNING
Wear safety glasses while test device is connected to the test port.
&.
CAU
Tl
ON
Be careful notto plugthe outputofthe primary regulatorwithout having
a
pressu 陀 relief
valve in the output circui
t.
1. Setthe system switch to Standby.
2. Disconnect all pipeline supplies.
3. Remove the
upper 陀ar
panel (Section 9.2).
For 3rd gas cylinder
regulato 邸,
also remove the lower rear pane l.
4. If equipped
,
turn the auxiliary O
2
flowmeter control fully clockwise (no flow).
5. Install a full cylinder in the cylinder supply to be tested. It is essential that
the cylinder be within 10% of its full pressure.
6. Remove the plug from the test port and connect a test device capable of
measuring 689 kPa (100 psi).
08/07 Ml110140