GE Aespire 7100 Technical Reference Manual page 148

Anestesia machine
Table of Contents

Advertisement

Aespire 7100
5-16
13. Carefully pull the collarbackso there is a slightgap between collar and the
valve body (but still engages the valve stop).
Collar
5e除crew
(example)
...... •
14. Tighten the collar setscrews. Start with the one opposite the tab if
possible.
15. Turn the valve stem counterclockwise at least one revolution to make sure
the collar tab clears the valve stop.
I
5tops must clear
If stops do not clear:
a. Turn the valve stem back to minimum position.
b. Loosen the collar setscrews.
c. Repeat steps 11 through 14.
16. Turn the valve stem clockwise to the minimum stop.
17. Verify there is no flow at the end of the tube.
18. Th oroughly clean the end of the nylon tube and reconnect it to the
vaporizer manifold inle
t.
19. Set the knob so that at minimum flow the label text is horizontal and the
knob is on an even plane (front to back) with the N
2
0 and O
2
knobs.
20. Replace the flowmeter panel shield and the rear pane l.
08/07 Mll10140