Generaltroubleshooting - GE Aespire 7100 Technical Reference Manual

Anestesia machine
Table of Contents

Advertisement

Aespire
7100
7.1 General Troubleshooting
WARNING
Objects in the breathing system can stop gas
f1
0w
to the patient
Th
is can cause
injury or death:
problem
High Pressure Leak
Do not use a test plug that is small enough to fall into the breathing system.
Make sure that there are no test plugs or other objects caught in the
breathing system.
Posslble Cause
Pipeline leak
O
2
flush valve
System switch
Cylinder not installed properly
Cylindergauges
Cylinder gaskets
Reliefvalves
Actlon
Use a leak
detecωr
or Snoop to check for source of leak.
Repair or replace defective
pa 内S.
Use a leak detector or Snoop to check for source of leak.
Make sure tubing connections are tigh
t.
Replace valve if defective.
Use a leak detector orSnoop to check for source ofleak.
Make sure tubing connections are tigh
t.
Replace switch if defective.
Make sure cylinder is correctly aligned.
V'erify that tee handles are tigh
t.
Use a leak detector or Snoop to check for source of leak.
Replace gauge if defective.
Use a leak detector or Snoop to check for source of leak.
Replace gasket if
de币ective.
Use a leak detector or Snoop to check for source of leak.
Replace valve if defective.
Low Pressure Leak
Vaporizer not installed properly
Reseat vaporizer if not installed
prope 川y.
(with vaporizer mounted)
7-2
Have vaporizer
servic坦d
at vaporizer center if vaporizer
leaks.
Missing or damaged o-ring on vaporizer
Check condition of o-ring.
manifold
Replace if missing or damaged.
Loosefill
po此
Check
fill
po凡 Tighten
if loose.
08/07 Mll10140