GE Aespire 7100 Technical Reference Manual page 202

Anestesia machine
Table of Contents

Advertisement

Aespire
7100
Message
Alarmj
Err
or
Cause
User ActlonjConcems
Servlce Repalr
No
Ex
p Flow Sensor
UserAlarm
Electrical signals show the
Connect the flow sensors.
flow sensor is not
Make sure the flow sensor
connected.
module is on all the way.
No Insp Flow Sensor User Alarm
Electrical signals show the
Connect the flow sensors.
flow sensor is not
Make sure the flow sensor
connected.
module is on all the way.
No O
2
Pressure
UserAlarm
The O
2
supply has failed.
Airflowwill continue. Ventilate
manually if
necessar如
Connect a pipeline supply or
install an O
2
cylinder.
O
2
Flush Failure
UserAlarm
The pressure switch that
Th is alarm occurs if you hold
If persists
,
check:
detects flush flow has
down the Flush button for
• O
2
flush switch
seen a very long flush
more than 30 seconds.
• Cable to SICB
(注30
sec).
• SICB
• Cable
ωCB/J11
O
2
Sensor Out of Circ User Alarm
O
2
sensor not installed in
Install a breathing circuit module
breathing circuit module.
and an O
2
sensor.
Sensor not measuring gas in
breathing circui
t.
O
2
Mon Disabled
UserAlarm
An Oxygen cell has been
Th e Oxygen Monitoring feature is Service Mode:
connected to a non-
not active on this system.
• System
active ventilator
Con自 gurations
monitoring feature.
On Battery-
UserAlarm
The mains supply is not
Ventilate manually to save
PowerOK?
connected or has failed
power.
and the system is using
At full charge
,
the
ba忧ery
ba吐ery
power.
permits approximately
30 minutes of mechanical
ventilation.
Make sure power is connected
and circuit breakers are
closed. Check the ventilator
fuse.
Patient Circuit Leak? User Alarm
Ex
haled volume <50% of
Check breathing circuit and flow
inspired volume for at
sensor connections.
least 30 seconds
Patient circuit leak audio can be
(mechanical ventilation).
turned off in the Alarm
Settings menu.
7-36
08/07 Mll10140