Aespire 7100
4-
22
Real-nme Value
7. To start calibration select MStart Pressure Sensitivity
Cal飞
•
Adjustthe APLand fresh gas flow until the real-time pressure reads a stable 10
cm H
2
0 - after 5 seconds
,
select MSave
Value飞
•
Adjustthe APL and fresh gas flow until the real-time pressure reads a stable 20
cm H
2
0 - after 5 seconds , select MSave Value".
•
Adjust the APLand fresh gas flow until the real-time pressure reads a stable 40
cm H
2
0 - after 5 seconds , select .Save Value".
•
Adjustthe APLand fresh gas flow until the real-time pressure reads a stable 60
cm H
2
0 - after 5 seconds , select MSave Value".
08/07
Ml110140