Page 67
User must manually accept results and auto accept results radio buttons
whether the outcome of the test are manually or automatically accepted. You may select user
must manually accept results (default) or auto accept results and return online.
• If user must manually accept results is selected, the results of the calibration will not be
• If auto accept results and return online is selected, the slope and R
Return to online mode checkbox
following the TOC calibration. You may enable or disable (default) this option. When enabled,
the analyzer returns to online mode following the TOC calibration. When disabled, the
analyzer stays at the bottle mode dialog box. All valves are closed, the pump is off, and the
lamp is off.
Slope limit text box
to pass. You may enter a value from 1 to 15%. The default value is 10%.
R2 limit text box
You may enter a value from 0.9000 to 0.9999. The default value is 0.9900.
Return icon
noted in the audit trail.

Conductivity calibration tab

User must manually accept results and auto accept results radio buttons
whether the outcome of the test are manually or automatically accepted. You may select user
must manually accept results (default) or auto accept results and return online.
• If user must manually accept results is selected, the results of the calibration will not be
• If auto accept results and return online is selected, the slope limit options are enabled.
Operator Manual
used until you accept them.
enabled. If the calibration is accepted, the new calibration factors are used.
allows you to specify the acceptable limit of the slope for the calibration
allows you to specify the acceptable limit for the determination coefficient.
returns you to the setup dialog box and saves any changes. Any changes are
Figure 5-21 Bottle test setup dialog box, conductivity calibration tab
used until you accept them.
If the calibration is auto accepted, the new calibration factors are used.
Setup – PAT700 Total Organic Carbon Analyzer
tells the analyzer whether or not to return to online mode
determine
2
limit options are
determine
Anatel