Use An Sd Card; Calibration - Hach 5500sc PO4 3-HR Operations

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 3

Use an SD card

Use an SD memory card to update the software and firmware and to download the event and data
logs. The SD icon is visible in the upper status bar of the main measurement screen when a card is
installed. The manufacturer recommends to use an SD card with a minimum of 2 GB storage
capacity.
1. Install the SD card (refer to
2. Select SD CARD SETUP from the MAIN MENU.
Note: The SD CARD SETUP option shows only when an SD card is installed.
3. Select an option.
Option
UPGRADE SOFTWARE
(conditional)
SAVE LOGS
MANAGE
CONFIGURATION
WORK WITH DEVICES
Update the firmware
Use an SD card with an upgrade file to update the firmware for the controller, sensor or network card.
The upgrade menu is shown only when the SD card contains an upgrade file.
1. Install the SD card into the SD card slot.
2. Select SD CARD SETUP from the MAIN MENU.
Note: The SD CARD SETUP option shows only when an SD card is installed.
3. Select UPGRADE SOFTWARE and confirm. Select the device and upgrade version, if applicable.
4. When the upgrade is complete, the display shows TRANSFER COMPLETE. Remove the SD
card.
5. Restart the instrument for the upgrade to take effect.

Calibration

The manufacturer recommends calibration of the analyzer after 1 day of operation to allow all system components
to stabilize.
The automatic calibration uses the installed known standard to calibrate the analyzer. Calibration can
be done manually or scheduled to occur automatically. Use the calibration menu to view calibration
data, start or cancel the auto calibration, override the auto cal settings, calibrate manually or to reset
to the default calibration.
1. Push cal to display the calibration menu.
Option
START MANUAL CAL
SET AUTO CALIBRATION Refer to
Figure 4
on page 6).
Description
Shows when an the upgrade file is available on the SD card. Select the specific
device for the upgrade.
Selects the device for the download data and to save the logs for last day, last
week, last month or all.
Saves and restores the backup settings, restore settings and/or transfer
settings between instruments.
READ DEVICE FILES—selects the data for each device to save on the SD
card. Options: sensor diag, measurement data (curve data for a measurement
cycle), cal history, cal data and/or test script. WRITE DEVICE FILES—shows
when an upgrade file is available for a new measurement cycle script.
N O T I C E
Description
Refer to
Start a calibration manually
Schedule automatic calibrations
on page 18.
on page 18.
English 17

Advertisement

Table of Contents
loading

Table of Contents