PAT700 Total Organic Carbon Analyzer – Installation

Chapter 2 Installation

DANGER
Tasks involving installation of this equipment have fire, electrical, and pressure related
hazards associated with them. These tasks must only be attempted by individuals
trained and knowledgeable in the particular task and the associated hazards.
2.1

Installation requirements

The quick-connect version of the PAT700 has an IP46 enclosure surrounding the electronics
portion of the analyzer. The conduit version of the PAT700 has a waterproof IP56 enclosure,
that can withstand water jets, surrounding the electronics portion of the analyzer. The
hydraulics portion of the analyzer is protected by a splash resistant enclosure, but does not
carry an IP rating.
• Install the analyzer in a dry, dust-free environment.
• Install the analyzer in a properly ventilated location where the ambient air temperature
• The analyzer has 3/4-inch NPT conduit fittings or quick-connect fittings for wiring.
See
The analyzer ship kit includes the following parts:
• This instruction manual
• 5-foot (1.5 meter) PFA inlet tubing
• 10-foot (3 meter) polypropylene outlet tubing
• 5/16-inch T-handle Allen wrench
• Stylus
• Instrument screwdriver
• USB memory stick
For units with quick connect I/O fittings the following items are also included:
• Quick connect power cord
• Male 8-pin quick connect cord fitting
• Three male 5-pin quick connect cord fittings
2.2

Orientation and mounting

NOTE
Failure to mount the analyzer in an upright position can cause measurement error.
To ensure proper operation of the measurement cell between analysis cycles, mount the
analyzer in an upright position. See
• Secure the analyzer using the factory-supplied mounting bracket (see
Anatel
is 10 to 40 °C (50 to 104 °F).
Figure 2-1, page 10
for mounting dimensions. The analyzer weighs 30 lb (13.6 kg).
page
12). Use the mounting holes in the bracket to mount it to a flat, stable surface.