Serial Ports - Hach ANATEL PAT700 Operator's Manual

On-line total organic carbon analyzer
Hide thumbs Also See for ANATEL PAT700:
Table of Contents

Advertisement

PAT700 Total Organic Carbon Analyzer – Installation
2.6

Serial ports

The analyzer has the following serial ports:
• A female RS-232 serial port for data acquisition and control via a host computer.
• A male RS-232 serial port for connection to an external printer (9600 baud, 8 data bits,
• A USB port for data transfer via a memory stick.
• An Ethernet
• All serial ports are internally wired.
Table 2-11
Description
Female RS-232 data acquisition port
Male RS-232 printer port
USB host
Ethernet
Figure 2-11
Anatel
1 stop bit, no parity).
®
10/100 base-T interface port for communication via Modbus
For information about using Modbus protocol, see the instruction manual entitled Using
Modbus Protocol with the Anatel PAT700 TOC Analyzer.
communication.
Serial communication specifications
®
interface
illustrates the RS-232 interface 9-pin connector.
Figure 2-11 RS-232 interface 9-pin connector
Table 2-11
lists specifications for serial
Specification
ASCII, 8 data bits, no parity, 1 stop bit, 9600 baud
8 data bits, no parity, 1 stop bit, 9600 baud
Output to USB flash drive
10/100 Mbits/sec., Modbus
CTS
Page 25
®
protocol.
®
TCP/IP
Operator Manual

Advertisement

Table of Contents
loading

Table of Contents