In conductivity mode, the analyzer starts by verifying that the UV lamp is off and the cell valve
is open to flush the cell. The analyzer reports the current readings for conductivity and
temperature every time a new value is available. The reported value is an average of the data
over the last two seconds. The average reading is also sent to the printer, based on the
printout settings, and to the data log. One data point is written to the log every 60 seconds.
4.5

Standby mode

In standby mode, the analyzer verifies that the lamp is off and the valves are open. The
valves, lamp, and pump states are based on the flow during idle setting. (See
page
4.6

Offline mode

In offline mode, the analyzer verifies that the lamp is off, the valves are closed, and flow
through the analyzer has stopped. You can terminate offline mode by switching to another
mode.
Operator Manual
49.) You can terminate standby mode by switching to another mode.
Figure 4-5 Home screen in standby mode
Figure 4-6 Home screen in offline mode
Run Modes – PAT700 Total Organic Carbon Analyzer