PAT700 Total Organic Carbon Analyzer – Diagnostics
Chapter 8 Diagnostics
8.1

Accessing the diagnostics dialog box

• At the home screen, touch the << icon to open the sliding toolbar, then touch the
• Touch the dual UV lamp status indicator in the lower right corner of the screen.
8.2

Diagnostics dialog box

The diagnostics dialog box provides access to lamp management, system tests, and
calibration dates.
8.3

General tab

The general tab displays the operating status of lamps 1 and 2. See
Version information text label
information.
Status icon
use, the icon is dimmed.
Hours left text label
should replace the UV lamp after 4300 hours of operation.
Anatel
diagnostics icon, or
Figure 8-1 Diagnostics dialog box, general tab
represents the status of each of lamp 1 and lamp 2. If a lamp is not currently in
Good
(green) means the lamp has not reached 0 hours left and is producing a
sufficient level of UV light based on the UV Detect™ diagnostics.
Marginal
(yellow) means the lamp has either reached 0 hours left or has been
determined by UV Detect diagnostics to be producing a diminished level of UV light
and that the lamp should be replaced.
Failed
(red) means the lamp has failed and is not longer producing sufficient UV light
to perform full sample oxidation.
Not in use
(dimmed) means the lamp is not in use.
counts down the number of hours remaining for the UV lamp. You
displays the firmware and operating system version
Page 111
Figure
8-1.
Operator Manual