Sampling With Pressure; Sampling From A Zero Pressure System; Conductivity Mode - Hach ANATEL PAT700 Operator's Manual

On-line total organic carbon analyzer
Hide thumbs Also See for ANATEL PAT700:
Table of Contents

Advertisement

PAT700 Total Organic Carbon Analyzer – Run Modes

Sampling with pressure

CAUTION
Potential burst hazard! Do not exceed the maximum 100 psi (690k Pa) sample pressure
ratings.
NOTE
Using the sample pump on a pressurized system could damage the pump. Do not use
the sample pump on a pressurized system.
In online TOC mode, the cell valve opens to flush the cell.
• When flushing is completed, the cell's sample valve closes, the analysis begins, and
• When oxidation is complete, the analyzer reports the results to the home screen, the

Sampling from a zero pressure system

If you are sampling from a zero pressure system, you can enable the pump option to draw
water into the cell. For instructions on setting up the pump option, see TOC,
page
4.4

Conductivity mode

In conductivity mode, the home screen displays conductivity and temperature readings. See
Figure
Anatel
initial conductivity and temperature readings are fed into the TOC algorithm.
Conductivity is monitored until oxidation is complete.
data log, and the printer, based on the printer settings and if the printer is connected.
The analyzer then checks to see if the elapsed time is greater than or equal to the
value for cycle time in settings. If it is, the next online TOC analysis is started.
Otherwise, the analyzer enters the idle mode.
48.
4-4.
Figure 4-4 Home screen, current tab, conductivity mode
Page 43
General tab,
Operator Manual

Advertisement

Table of Contents
loading

Table of Contents