Log View Tab - Hach ANATEL PAT700 Operator's Manual

On-line total organic carbon analyzer
Hide thumbs Also See for ANATEL PAT700:
Table of Contents

Advertisement

PAT700 Total Organic Carbon Analyzer – Startup
Date and time
the current time format setting, the date and time is in 24-hour format (yyyy-mm-dd hh:mm:ss)
or 12-hour format (yyyy-mm-dd hh:mm:ss AM/PM). The date and time are configured in the
setup dialog box.
Bottles icon
bay. There are three states for this indicator (no bottles loaded, bottles loaded, or bottles
empty). Touching the icon takes you to the bottle mode dialog box.
Status indicator
• Green indicates the UV lamp is good.
• Yellow indicates the UV lamp is marginal and should be replaced.
• Red indicates the UV lamp has failed and must be replaced.
Process animation
animated graphic. There are five possible states: water flowing through the cell, water not
flowing through the cell, analyzing, new alarm condition, and acknowledged alarm condition.
Analysis results are displayed until a new analysis is complete. It displays the last readings for
conductivity, temperature, and the time of the last reading.

Log view tab

The log view tab displays all analysis readings or events as they would appear on the printout.
The display include all data, up to the last 72 hours, since the last power-up. See
Anatel
Temperature
is the last temperature reading in "Temperature: X.X °C" format or
"Temperature: X.X °F" format, depending on the temperature units selected in the
setup dialog box.
Sample time
is the time of the start of the last reading in 12-hour "Sample Time:
hh:mm:ss" format.
Trend
is the TOC trend over the last hour in "Trend: +/-X.X ppb/hr" format.
Curve type
is the curve type for the last TOC reading in "Curve Type: X" format, where
X can be one of four types (1, 2, 3, or 5).
(top of screen) displays the current date and time in ISO format. Depending on
(lower left corner of screen) displays the current state of the bottles in the bottle
(lower right corner of screen) displays the status of the two UV lamps.
(right side of screen) displays the current state of the analysis as an
Figure 3-4 Home screen, log view tab
Page 39
Figure
3-4.
Operator Manual