Page 96
Filename text box
Exporting record no: X bar graph indicator
Export icon
to export, that you have entered a valid file name, and that a USB flash drive is inserted into
the analyzer's USB port. The file is written to the root directory of the flash drive. If the file
already exists, you are asked if you want to overwrite it.
Return icon
6.10

Excursion mode

NOTE
Excursion mode requires the use of an RFID tagged excursion bottle. Excursion mode
does not operate without an RFID tagged excursion bottle.
Excursion mode enables the analyzer to withdraw a sample from the water system when a
predefined TOC level is exceeded or when alarm condition 35 or 36 occurs. An excursion
sample can also be implemented manually. With excursion mode enabled and an empty
excursion sample bottle loaded, a sample of the process system water is drawn immediately
following the indication of the appropriate event. The water sample is then available for
analysis to help determine the cause of the water system excursion. Excursion sample bottles
must contain an RFID tag. The information from the excursion event is written to the bottle's
RFID tag to ensure accurate information about the water sample.
Excursion with validation includes the use of a 500 ppb sucrose standard with the excursion
sample bottle. The 500 ppb standard is used for verifying the performance of the analyzer
immediately following the excursion event and sample. If excursion with validation is enabled,
the analyzer draws a water sample, immediately following the analysis that indicated an
excursion, followed by an analysis of the 500 ppb standard.
Touch the excursion mode button in the bottle mode dialog box. See
excursion mode dialog box appears, as illustrated in
Operator Manual
Figure 6-23 Export grab sample dialog box
allows you to specify the filename for the exported data file.
starts the export. Before the export begins, the analyzer verifies that there is data
returns you to the security dialog box.
Bottle Mode – PAT700 Total Organic Carbon Analyzer
displays the progress of the export.
Figure 6-1, page
Figure
6-24.
76. The
Anatel