PAT700 Total Organic Carbon Analyzer – Bottle Mode
6.8

System suitability test

CAUTION
Potential puncture hazard! Contact with the exposed needle could cause serious
injury. Use extreme care when installing and removing the sample bottles.
A system suitability test allows you to test analyzer measurements according to the guidelines
established in USP Method <643> using 500 ppb sucrose and 500 ppb 1,4-benzoquinone
standards.
1)
2)
3)
4)
5)
6)
Anatel
Allow the system suitability test standards to warm to room temperature.
At the home screen, touch the << icon to open the sliding toolbar, then touch the
bottles icon in the toolbar, or touch the bottles icon in the lower left corner of the
screen.
Touch the run standards button in the bottle mode dialog box. See
page
76.
Touch the system suitability button in the run standards dialog box. See
page
77.
Touch the forward icon. The OASIS door will unlock for five seconds. If the door is
not opened within five seconds, the unlock icon can be used to unlock the door
manually.
Install the standards bottles following the instructions as shown in the load bottles
dialog box and ensuring that you place the standards bottles in the correct
positions. See
Figure
• Bottle 1 = blank
• Bottle 2 = empty
• Bottle 3 = 500 ppb sucrose
• Bottle 4 = 500 ppb benzoquinone
Figure 6-18 Load bottles dialog box, system suitability test
6-18. From left:
Page 91
Figure 6-1,
Figure 6-2,