Page 49
• A high to low state transition on digital1 aborts the current operation and immediately
• If digital2 is high, the analyzer enters online TOC analysis mode. If digital2 is low, the
Average last results checkbox
displayed on the home screen and in the log view, logged to the data log, and sent to the
printer based on the printer settings. The default value is disabled.
Return icon
noted in the audit trail. All changes take effect on the next change of state.

Idle tab

Flow during idle radio buttons
while waiting for the next analysis to begin if cycle time is greater than the time required for
cell flush and oxidation. You may select without conductivity, with conductivity (default), or
none.
• If without conductivity is selected, the sample valve opens during idle, allowing
• If with conductivity is selected, the sample valve opens during idle, allowing continuous
• If none, the analyzer sample valve closes, stopping all flow through the sample cell
Return icon
noted in the audit trail. All changes take effect on the next change of state.
Operator Manual
runs a single TOC analysis. The analysis runs to completion regardless of whether
subsequent trigger values are received. Once the analysis is complete, the analyzer
returns to the mode defined by digital2.
analyzer enters conductivity mode.
returns you to the setup dialog box and saves the changes. Any changes are
continuous flow through the cell. No conductivity is reported.
flow through the cell. Conductivity of the water is reported.
during idle.
returns you to the setup dialog box and saves the changes. Any changes are
averages the last X TOC analyses. The averaged value is
Figure 5-3 TOC dialog box, idle tab
allows you to select the state that the analyzer should enter
Setup – PAT700 Total Organic Carbon Analyzer