Parameter Setup Command - Hach ANATEL PAT700 Operator's Manual

On-line total organic carbon analyzer
Hide thumbs Also See for ANATEL PAT700:
Table of Contents

Advertisement

Page 30

Parameter setup command

Set/get system date and time
"SY"
Example:
Read command:
Response:
"7
Write format:
"SY
"SY
Disable automatic result
"CA"
printouts for serial port
Set/get global settings
"HR"
1. Flush time (HH:mm:SS)
2. Cycle time (HH:mm:SS)
3. TOC alarm limit (ppb)
4. Pump sampling (0 = Disabled, 1 = Enabled)
5. Sensor name (1 to 13 characters)
6. Idle mode (0 = No flow, 1 = Flow with conductivity, 2 = Flow without conductivity)
7. Always 0
8. Resistivity/conductivity units (0 = Mohm-cm, 1= μS/cm C, 2 = Mohm-cm U, 3 = μS U)
9. Always 0
10. TOC mode print strategy (0 = Continuous, 1 = Percent change, 2 = Disabled)
11. Conductivity print strategy (0 = Timed, 1 = Percent change, 2 = Disabled)
12. Active alarm type (0 = TOC only, 2 = TOC and uncompensated conductivity)
13. Analog output type (0 for 4-20 mA, 1 for 0-20 mA)
14. Analog output on error (0 = Minimum [2mA], 1 = Last output, 2 = Maximum [22mA])
15. TOC percent change between printouts (1-99)
16. Conductivity percent change between printouts (1-99)
17. Analog zero-scale TOC (ppb)
18. Analog full-scale TOC (ppb)
19. User TOC calibration slope
20. Digital control mode (0 = Disabled, 1 = Enabled)
21. Always 0 (data and audit Log)
22. User conductivity calibration slope
Operator Manual
"SY"
25
2007
20
33
mm
DD
YYYY
HH
7
30
2007
12
Disables the automatic printout of results for the serial port.
Sets or gets global settings. Compatible with or without user
conductivity calibration slope parameter.
Installation – PAT700 Total Organic Carbon Analyzer
31"
MM
SS"
30
25"
Anatel

Advertisement