AVEA ventilator systems
Chapter 7: Alarms and Indicators
Status Indicators
The ventilator displays the following status indicators.
Compressor Active
If the internal compressor is active, the Compressor Active icon shown here will display at the bottom of the touch
screen with no accompanying tone.
Heliox Source Connected
If Heliox gas is connected this green icon displays in bottom right of the touch screen.
Mains/Battery Indicators
There are visual status indicators on the ventilator front panel for the mains power and the internal and external batteries
(Figure 7–1).
The sequence in which the power sources are used by the ventilator is:
Mains AC Power
•
External Battery (if installed)
•
Internal Battery
•
Power On Indicator
The green Power On indicator lights up whenever the power switch is on ( I ) and power is being supplied from any of the
available power sources (AC, external battery, or internal battery).
On battery indicator while operating on internal or external battery, a battery icon will blink in the lower right hand corner
of the display.
AC Power Indicator
The green AC indicator is on whenever the ventilator is connected to AC power. It displays whether the power switch
is on ( I ) or off (O).
Operating On Battery Indicator
When operating on battery power (Internal or External) a yellow blinking battery indicator will appear in the lower right
hand corner of the LCD screen.
External Battery Power Indicator
The EXT indicator above the battery status indicators is lit whenever the external battery is providing the primary source
of power for the ventilator.
L2786 Rev. M