Alarms Suspended During Ncpap With Rate Set To Off - CareFusion AVEA Operating Instructions Manual

Hide thumbs Also See for AVEA:
Table of Contents

Advertisement

AVEA ventilator systems

Alarms suspended during nCPAP with Rate set to off

Time-based alarms
High Rate
I-Time Limit
I:E Limit
Apnea Interval
Low Ppeak
Range: 1 to 40 cmH
Default: 7 cmH
Note:
Not available in nasal CPAP mode unless Rate of one or greater is set.
Alarm threshold setting changes during nCPAP / IMV will not be retained for other ventilation modes following exiting
from nasal CPAP mode.
High Ppeak
Range: 2 to 45 cmH
Default: 20 cmH
Note:
Not available in nasal CPAP mode unless a Rate of one or greater is set.
Alarm threshold setting changes during nCPAP / IMV will not be retained for other ventilation modes following exiting
from nasal CPAP mode.
L2786 Rev. M
Chapter 6: Infant Non-invasive Ventilation
Volume-based alarms
High Ve
High Vt
Low Vte
Low Ve
Volume Limit
O (Neonate setting – nCPAP / IMV)
2
O
2
O (Neonate setting – nCPAP / IMV)
2
O
2
Pressure alarms
High Ppeak
Ext High Peak Alarm
Low PEEP
Low Ppeak
Occlusion
157

Advertisement

Table of Contents
loading