136
6.
Insert the airway adapter into the CO
Figure 5.5a Adult / Pediatric Adapter
Figure 5.5b Pediatric / Neonatal Adapter
Figure 5–5: Aiarway adaptors
7.
Perform the "sensor zero" procedure by following the instructions in the section "Zeroing the CAPNOSTAT 5" on page
144. The zeroing procedure must also be performed when switching between disposable and reusable airway adapters.
8.
After the sensor is successfully zeroed, place the airway adapter and sensor into the ventilator circuit between the wye
and endotracheal tube (and any adapters) as shown in the preceding illustration.
Chapter 5: Volumetric Capnography
sensor. The adapter clicks into place when properly inserted.
2
Operator's manual
L2786 Rev. M