Mip / 100 Pmaneuver Screen; Mip Maneuver Settings; Duration - CareFusion AVEA Operating Instructions Manual

Hide thumbs Also See for AVEA:
Table of Contents

Advertisement

AVEA ventilator systems
MIP / P
Maneuver Screen
100
Figure 4–12: MIP Maneuver Settings
The MIP (Maximum Inspiratory Pressure) / P
during the patient's active effort to demand a breath. During the maneuver, the inspiratory flow valve remains
closed and no inspiratory flow is delivered. The MIP is an indication of the maximum negative pressure that the
patient can draw, while P
breath.
Controls

Duration

L2786 Rev. M
Chapter 4: Monitors, Displays and Maneuvers
100
is an indication of the pressure drop that occurs during the first 100 milliseconds of the
100
The preset Duration shall determine the maximum amount of time that the maneuver will last.
Normal ventilation will be suspended for the duration of the maneuver and will resume after
the duration has timed out.
Range:
5.0 to 30.0 seconds
Default:
10 seconds
maneuver measures the negative deflection in the pressure tracing