Monitors; Ncpap Primary Controls And Alarm Threshold Indicators; Caution Message Display (Ncap) - CareFusion AVEA Operating Instructions Manual

Hide thumbs Also See for AVEA:
Table of Contents

Advertisement

AVEA ventilator systems
Note:
Low Nasal CPAP Pressure and High Nasal CPAP Pressure Alarm Thresholds are updated automatically when a new
value is accepted in the nCPAP Primary Control.
Figure 6–6: nCPAP Primary Controls and Alarm Threshold Indicators
CAUTION!
Apnea back-up ventilation is suspended during nCPAP.
AVEA continually displays the following message during nCPAP administration.
Figure 6–7: Caution Message Display

Monitors

In Nasal CPAP, all existing monitors will be suspended, except:
Air Inlet Pressure (Air Inlet)
Oxygen Inlet (O
Gas Composition Monitor (F
Percent Leak
The following monitors have been added for Nasal CPAP:
nCPAP level (mean airway pressure)
Range:
Resolution: 1 cmH
Accuracy:
CPAP Flow (mean inspiratory flow)
Range:
Resolution: 0.1 LPM
Accuracy:
L2786 Rev. M
Chapter 6: Infant Non-invasive Ventilation
Inlet)
2
)
IO
2
0 to 120 cmH
O
2
O
2
±3.5% of reading or ±2 cmH
0–300 LPM
±10%
O, whichever is greater
2
153