Apnea Backup in CPAP/PSV or APRV / BIPHASIC

When CPAP/PSV or APRV / BIPHASIC is selected, you MUST:
1.
Set the primary and advanced settings for CPAP/PSV or APRV / BIPHASIC.
2.
Select the breath type for APNEA backup mode (Volume or Pressure in adult and pediatric patients or Volume, Pressure
or TCPL in neonatal patients) by pressing the Apnea Settings key.
3.
Set the primary and advanced controls appearing at the bottom of the touch screen, for the selected apnea breath type
before pressing the MODE ACCEPT button. The controls for apnea backup ventilation will not be visible once the
MODE ACCEPT button has been pressed. Only the controls that are active and required for CPAP/PSV or APRV
/ BIPHASIC will remain.
See Figure 3–18 to Figure 3–21 for Apnea backup settings available in each mode.
Figure 3–18: Volume Apnea Backup settings for APRV / BIPHASIC Mode
Chapter 3: Ventilator Operation
Operator's manual
L2786 Rev. M