O2 Alarms - CareFusion AVEA Operating Instructions Manual

Hide thumbs Also See for AVEA:
Table of Contents

Advertisement

178
O
Alarms
2
Low O
% (Low F
2
This is a high priority audible/visual alarm. LOW F
Delivered O
% falls below the set F
2
High O
% (High F
2
This is a high priority audible/visual alarm. HIGH F
Delivered O
% rises above the set F
2
Table 7–1: Alarm Conditions
Message
SAFETY VALVE
Safety valve is open
OPEN
VENT INOP
Ventilator failure due to a recoverable or non-recoverable
condition. The safety valve opens, indicated by a SAFETY VALVE
message, and the patient is allowed to breathe room air. PEEP is
not maintained
LOSS, AIR
Wall air drops below 18.0 psig (1.2 bar) and no functional
compressor is installed or the compressor output is insufficient to
meet instrument demand. Patient will continue to be ventilated by
O
supply only.
2
LOSS, O
Oxygen supply to the ventilator drops below 18.0 psig (1.2 bar)
2
and the %O
by the air supply only
LOSS, HELIOX
The alarm is triggered if heliox is being used and the heliox gas
supply to the ventilator drops below 18.0 psig (1.2 bar). The
patient continues to be ventilated by the oxygen supply only.
LOSS, GAS
All sources of gas fail; wall air, internal compressor (if installed)
SUPPLY
and oxygen. The safety valve opens, indicated by a SAFETY
VALVE OPEN message, and the patient is allowed to breathe
room air. PEEP is not maintained.
NOT
The internal blended gas pressure is below 1 psi for greater than
VENTILATING
about 12 seconds, indicating there is no ventilation occurring.
The safety valve should be open in this condition, as there is no
pressure to keep the safety valve closed. This alarm is only a
monitor of this pressure level and does not change the "state" of
the ventilator. The condition is recoverable if pressure is restored.
LOW P
The peak inspiratory pressure for a breath is less than the set
PEAK
LOW P
HIGH P
Peak inspiratory pressure is greater than the set HIGH P
PEAK
Inspiration is terminated and the circuit pressure is allowed to
return to baseline pressure + 5 ± 1.5 cmH
breath is delivered.
EXT HIGH P
Activates whenever the HIGH P
PEAK
more than 5 seconds (i.e. If the circuit pressure fails to return to
PEEP + 5 cmH
Chapter 7: Alarms and Indicators
)
IO
2
minus 6% or 18% F
IO
2
)
IO
2
+ 6%.
IO
2
Alarm Condition
is set to > 21%. Patient will continue to be ventilated
2
Not active for spontaneous breaths.
PEAK.
alarm has been active for
PEAK
O within 5 seconds). The safety and exhalation
2
is displayed and a high priority tone sounds if the monitored
IO
2
whichever is greater.
IO
2,
is displayed and a high priority tone sounds if the monitored
IO
2
N/A
N/A
N/A
N/A
N/A
N/A
Pressure < 1psi for > 12 seconds
3 to 99 cmH
Default 3 cmH
Normal Breath Range:
PEAK.
Adult:
O before the next
10 to 105 cmH
2
Default: 40 cmH
Pediatric:
10 to 85 cmH
Default: 40 cmH
Neonate:
Default: 30 cmH
Sigh Breath Range:
1.5 x set normal HIGH P
Only active for sigh breaths
N/A
Operator's manual
Range
O
2
O
2
O
2
O
2
O
2
O
2
O
2
PEAK
L2786 Rev. M
Priority
High
High
High
High
High
High
High
High
High
High

Advertisement

Table of Contents
loading

Table of Contents