CareFusion AVEA Operating Instructions Manual page 228

Hide thumbs Also See for AVEA:
Table of Contents

Advertisement

214
DISPLAY
DESCRIPTION
Ti
Inspiratory time.
Te
Exhalation Time.
I:E
Inspiratory/expiratory ratio
Note: Not active for demand breaths.
Display
Description
f/Vt
Rapid shallow breathing index.
PRESSURE MONITORS
Ppeak
Peak inspiratory pressure.
Not active with spontaneous breaths
Pmean
Mean airway pressure.
Pplat
Plateau pressure. If no plateau occurs, then the
monitor displays * * *
PEEP
Positive end expiratory pressure.
Air Inlet
Air inlet gas supply pressure.
O
Inlet
Oxygen inlet gas supply pressure.
2
Pbaro
Barometric pressure
GAS COMPOSITION MONITORS
F
Delivered percent O
IO
2
MECHANICS
Cdyn
Dynamic Compliance (C
absolute and normalized to patient weight.
Cdyn/Kg
Cstat
Respiratory System Compliance (C
Static Compliance C
to patient weight.
Cstat/Kg
Note: This requires an Inspiratory Hold maneuver.
Rrs
Respiratory system resistance.
Note: Calculation is performed during an Inspiratory
Hold maneuver.
PIFR
Peak Inspiratory flow rate.
PEFR
Peak Expiratory flow rate.
Ccw
The ratio of the tidal volume (exhaled) to the Delta
Esophageal Pressure (dP
esophageal balloon.
Appendix D: Monitor Ranges and Accuracies
.
2
and C
/ Kg),
DYN
DYN
), (a.k.a.
RS
), absolute and normalized
STAT
). Requires an
ES
RANGE
ACCURACY
0.00 to 99.99
± 0.03 sec
sec
0.00 to 99.99
± 0.03 sec
sec
1:99.9 to
Derived from accuracies for monitored Ti
99.9:1
and Te
Range
Accuracy
0 to 500
Derived from accuracies for spontaneous
b
/min/L
breath rate and spontaneous minute
2
volume
0 to 120
± 3.5% of reading or ± 2 cmH
cmH
O
whichever is greater
2
0 to 120
± 3.5% of reading or ± 2 cmH
cmH
O
whichever is greater
2
0 to 120
± 3.5% of reading or ± 2 cmH
cmH
O
whichever is greater
2
0 to 60 cmH
O
± 3.5% of reading or ± 2 cmH
2
whichever is greater
0 to 80 psig
± 5 psig (1.4 – 5.5 bar)
0 to 80 psig
± 5 psig (1.4 - 5.5 bar)
760 to 545
± 2% of full scale
mmHg or 101
to 72.7 kPA
0 to 100%
± 3%
0 to 300
Derived
ml/cmH
O
2
0.00 to 5.00
ml/cmH
O⋅kg
2
0 to 300
Derived
ml/cmH
O
2
0.00 to 5.00
ml/cmH
O⋅kg
2
0 to 100
Derived
cmH
O/L/sec
2
0 to 300 L/min
± 10% of setting or ± (0.2 L/min + 10%
of setting), whichever is greater
(All patients)
0 to 300 L/min
± 10% of setting or ± (0.2 L/min + 10%
of setting), whichever is greater
(All patients)
0 to 300
+ 10%
mL/cmH
O
2
Operator's manual
O,
2
O,
2
O,
2
O,
2
L2786 Rev. M