14
4.
Drain Tube and Pinch Clamp. Inspect for any visible damage and make sure it is securely installed.
5.
Periodically inspect the filter-vial water level and empty it before it reaches the maximum-level line.
6.
To empty the fluid in the collection vial, press open the pinch clamp to empty contents of the collection vial into an
appropriate receptacle. Close and lock the pinch clamp when finished.
Figure 2–13: Draining water trap
WARNING
The drain tube must be fully attached to the filter and the pinch clamp must be in the closed position.
Figure 2–14: Closed pinch clamp
Figure 2–15: Open pinch clamp