208

VOXP Communications Requirements

Communication is established between a ventilator and an external system with a properly configured set of system-level
items from physical cables, adapters, and communication parameters to application protocols. A detailed description of
the interface is available in a separate specification, Consolidated VOXP Specification, part number L3058, Revision A,
or later.
To ensure proper operation of the ventilator, there is an important limitation to the implementation of the interface with
external devices. Waveforms should not be selected at baud rate settings less than 57600, and no more than 3
waveforms should be selected at baud rates of 57600 and higher. Waveform transmission requires high speed
communications, and problems can occur if waveforms are requested at lower baud rates, or if too many waveforms are
requested at high baud rates.
WARNING!
To ensure proper operation of the ventilator, external device communications using waveforms must follow
the recommendations in Consolidated VOXP Specification, part number L3058, Revision A, or later. The MIB
port must be connected to a device that meets the IEC60601–1 standard.
Printer
The ventilator has a standard 25-pin female Centronics parallel printer port for interfacing to an external printer.

Remote Nurse Call

The ventilator has a modular jack configured to interface with external systems that are either wired for normally open
(N.O., close on alarm) or normally closed (N.C., open on alarm) signals.
In the active state, the remote alarm can sink 1.0 A.
Video Output
The ventilator provides a video output connector, which allows for interfacing to an externally located 256-color, 800 x
600, SVGA monitor. The Video Output is always enabled.
Language Support
The list of selectable languages on the AVEA are English, Chinese, Czech, Dutch, French, German, Greek, Hungarian, Italian,
Japanese, Polish, Portuguese, Russian, Spanish, and Turkish.
Atmospheric & Environmental Specifications
Temperature and Humidity
Storage
Temperature:
Humidity:
Operating
Temperature:
Appendix B: Specifications
−20 to 60º C (−4 to 140º F)
0 to 95% RH non-condensing
5 to 40º C (41 to 104º F)