162
6.
Select Continue; the following message appears
7.
If characterization is successful the following message appears
8.
Select Continue.
9.
Set the prescribed level for nCPAP Pressure and/or F
the desired value is displayed and by either touching the primary control again or by touching the ACCEPT
membrane key adjacent to the Data Dial to activate the new setting. If NIMV is prescribed, set the prescribed level
for Inspiratory Pressure, Inspiratory Time and Rate as well.
10. Connect the Nasal CPAP device to the patient.
Chapter 6: Infant Non-invasive Ventilation
by touching the primary control, turning the Data Dial until
IO
2
Operator's manual
L2786 Rev. M