Loops

Accessing the Loops Screen

To access the loops screen press the screens
membrane button to the left of the touch screen on
the UIM. The button is labeled with the icons shown
here.
International
Select LOOP from the options that appear.

Choice of Loops

The ventilator displays 2 loops in real time, selected from the following.
Vt-Flow
PAW - Vt
PES - Vt
PTR - Vt
P
– Vt
INSP
P
– Vt
Tp
PCO
/Vte
2
Note:
Loops are circuit compliance compensated for pediatric and adult patient sizes.
Chapter 4: Monitors, Displays and Maneuvers
or
English
Flow / Volume Loop. Inspiratory flow / Volume. If proximal flow sensor is used values are
based on proximal flow sensor measurements.
Available for all patients.
Airway Pressure / Volume loop. Active for all patients.
Esophageal Pressure vs. Volume loop. This requires the use of an optional esophageal
catheter and is active for adult and pediatric patients only.
Tracheal Pressure vs. Volume loop. This requires the use of an optional tracheal catheter
and is active for adult and pediatric patients only.
Inspiratory Pressure vs. Volume loop.
Transpulmonary vs. Volume. This requires the use of an optional esophageal catheter and is
active for adult and pediatric patients only.
Exhaled CO
vs. Exhaled Vt
2
Figure 4–4: Screen Selection
Operator's manual
L2786 Rev. M