Selecting Language .

Touch the language box and use the data dial to select the desired language. Use the Accept key to accept the change.
All text displayed on the LCD screen will automatically be translated to the set language.
Note:
For ease of use all languages appear in their native text in the text selection box on the utilities screen.

Low Vte Alarm Sensitivity

Sets the number of consecutive breaths with an exhaled tidal volume below the Low Vte Alarm setting which are
required to sound the alarm. The default is 3 breaths; the range is 1-5 breaths.
Increase F
IO
2
Configures the step increase used during the increase oxygen maneuver. Sets the amount of oxygen the
ventilator will increase above the current set
Example:
The default setting for infants is 20% and 79% for Pediatric and Adult applications.
Note:
To achieve 100% delivered F
Note:
The settings will be reset to default values when New Patient is selected in the set up menu.
L2786 Rev. M
Chapter 2: Unpacking and Setup
F
IO
2
If the Increase
is set at 20%
F
IO
2
AND
The set
is 40%
F
IO
2
WHEN
The increase
Maneuver is activated the
F
IO
2
after which it will return to 40%.
during the Increase O
IO
2
.
F
IO
maneuver set the Increase F
2
will increase to 60% for two minutes
2
setting to its maximum of 79%.
IO
2