Volume Data; Breath Phase Data - CareFusion AVEA Operating Instructions Manual

Hide thumbs Also See for AVEA:
Table of Contents

Advertisement

204
Scale Factor:
Accuracy:
Zero Offset:
3.
Volume:
Range:
Scale Factor:
Accuracy:
Zero Offset:
4.
Breath Phase
The ventilator provides a continuous analog voltage representative of breath phase (Inspiration = 5 VDC, Expiration
= 0 VDC).
Appendix B: Specifications
1 L/min / 25 mV (Adult)
1 L/min / 50 mV (Pediatric)
1 L/min / 100 mV (Neonate)
± 10% of reading or ± 30 mV, whichever is greater
None
-1.00 to 4.00 L
(Adult)
-200 to 800 ml
(Pediatric)
-100 to 400 ml
(Neonate)
1 L / V
(Adult)
1 ml / 5 mV
(Pediatric)
1 ml / 10 mV
(Neonate)
± 10% of reading or ± 30 mV, whichever is greater
1.000 VDC
Operator's manual
L2786 Rev. M