Volumetric Capnography Specifications

Sensors
Sensor Type
Sensor Physical Characteristics
Sensor Compatibility
CO
Measurement
2
CO
Measurement range
2
CO
Measurement Accuracy
2
CO
Resolution
2
CO
Stability
2
Gas Composition Compensation
Airway Adapters
Adult/Pediatric
Single Patient Use
Infant /Pediatric
Single Patient Use
Adult/Pediatric
Reusable
Infant /Pediatric
Reusable
All components are Latex free
Appendix E: Sensor Specifications & Circuit Resistance
Mainstream, non-dispersive infrared single-beam optics, dual wavelengths. No
moving parts
Weight: 25 g (78 g with standard cable and connectors)
Size: 33 mm x 43 mm x 23 mm. Cable length: 3 m
The CareFusion Capnostat 5 is interchangeable between CareFusion equipment
only.
0 – 150 mmHg (0 – 20 kPa)
± 2 mmHg for 0-40 mmHg
± 5% of reading for 41-70 mmHg
± 8% of reading for 71-100 mmHg
± 10% of reading for 101-150 mmHg
1 mmHg
< 0.8 mmHg over four hours
Oxygen and Helium gas composition. Automatic compensation
For use with endotracheal tube greater than 4mm ID
Dead space: 5 mL
Weight: 7.7 g
Color: Clear
For use with endotracheal tube less than or equal to 4mm ID
Dead space: < 1 mL
Weight: 9.1 g
Color: Purple
For use with endotracheal tube greater than 4mm ID
Dead space: 5 mL
Weight: 12 g
Color: Black
For use with endotracheal tube less than or equal to 4mm ID
Dead space: < 1 mL
Weight: 14.9 g
Color: Red