CareFusion AVEA Operating Instructions Manual page 194

Hide thumbs Also See for AVEA:
Table of Contents

Advertisement

180
Message
becomes disconnected from the slave ventilator during ILV.
INVALID GAS ID
A medium priority audible/visual alarm shall be activated, and
INVALID GAS I.D. shall be indicated whenever a defective gas
I.D. connector is installed in the ventilator. When a defective Gas
I.D. connector is detected, the gas corrections default to air.
FAN FAILURE
A low priority audible/visual alarm is activated, and FAN FAILURE
indicated, whenever the fan has stopped rotating.
High EtCO
Low priority alarm if monitored EtCO
2
threshold setting.
Low EtCO
Low priority alarm if monitored EtCO2 does not exceed the alarm
2
threshold setting.
Chapter 7: Alarms and Indicators
Alarm Condition
exceeds the alarm
2
Range
N/A
N/A
6 – 150 mmHg or off
Default: 60 mmHg
1 – 145 mmHg or off
Default: 30 mmHg
Operator's manual
Priority
Medium
Low
Low
Low
L2786 Rev. M