Primary Breath Controls

The Primary Breath Controls are the operator set controls, which directly affect the way a breath is delivered to your
patient. They are displayed along the bottom of the AVEA LCD touch screen. Only the active controls for the selected
mode of ventilation will be displayed.
Table 3–3: Primary Breath Controls
Displayed
Description
Control
bpm
Breath rate shown in Breaths
Rate
per Minute
ml
Tidal Volume in milliliters
Volume
cmH
O
Inspiratory Pressure in
2
centimeters of water pressure
Insp Pres
L/min
Peak Inspiratory Flow in Liters
Peak Flow
per Minute
sec
Inspiratory Time in Seconds
Insp Time
sec
Sets an inspiratory pause which
Insp Pause
will be in effect for each Volume
breath delivered
cmH
O
Pressure Support in centimeters
2
PSV
of water pressure
cmH
O
Positive end expiratory pressure
2
PEEP
in centimeters of water pressure
L/min
Sets inspiratory flow trigger
Flow Trig
point in liters per minute
%
Controls the percentage of
%O
oxygen in the delivered gas.
2
cmH
O
In APRV / BIPHASIC mode,
2
Pres High
controls the baseline pressure
achieved during Time High.
sec
In APRV / BIPHASIC mode sets
Time High
the minimum time for which the
high-pressure setting is
maintained.
sec
In APRV / BIPHASIC mode sets
Time Low
the minimum time for which the
low pressure setting is
maintained.
cmH
O
In APRV / BIPHASIC mode
2
Pres Low
controls the baseline pressure
achieved during Time Low.
Chapter 3: Ventilator Operation
Range
1 to 150 bpm (Neo / Pediatric)
1 to 120 bpm (Adult)
0.10 to 2.50 L (Adult)
25 to 500 ml (Pediatric)
2.0 to 300 ml (Neonate)
0 to 90 cmH
O (Adult/Pediatric)
2
0 to 80 cmH
O (Neonate)
2
3 to 150 L/min (Adult)
1 to 75 L/min (Pediatric)
0.4 to 30.0 L/min (Neonate)
0.20 to 5.00 sec
(Adult/Pediatric)
0.15 to 3.00 sec (Neonate)
0.0 to 3.0 sec
0 to 90 cmH
O (Adult/Pediatric)
2
0 to 80 cmH
O (Neonate)
2
0 to 50 cmH
O
2
0.1 to 20.0 L/min
21% to 100%
0 to 90 cm H
O
2
0.20 to 30.0 sec
0.20 to 30.0 sec
0 to 45 cmH
O
2
Accuracy
± 1 bpm
± (0.2 ml + 10% of setting)
Adult/Pediatric: ±4cmH
O +5%
2
Neonate: ±3 cmH
O +2.5%
2
(measured at the patient wye, end inspiratory pressure
after 0.3 seconds)
± 10% of setting or ± (0.2 L/min + 10% of setting),
whichever is greater
± 0.10 sec
± 0.10 sec
± 3 cmH
O or ± 10% whichever is greater
2
± 2 cmH
O or ± 5% of setting, whichever is greater
2
+ 1.0 / − 2.0 L/min ( for PEEP < 30 cmH
+ 2.0 / − 3.0 L/min ( for PEEP > 30 cmH
± (0.2 L/min + 10% of setting) (Wye flow sensor only)
± 3% O
2
± 3 cmH
O
2
± 0.1 sec
± 0.1 sec
± 2 cmH
O or ± 5% of setting, whichever is greater
2
Operator's manual
O)
2
O)
2