Physiologic Dead Space / Tidal Volume Ratio (Vd / Vt phy)

Range: 0 – 99%
Resolution: 1%
Alveolar Dead Space (Vd alv)
Range: 0 – 999 mL
Resolution: 0.1 mL or three significant digits (whichever is greater)
Oxygenation Index (OI)
Oxygenation index is a dimensionless number often used to assess the "pressure cost" of oxygenation.
OI Range: 0 – 200 (when PAO2 is entered in mmHg); OI Range: 0 – 1500 (when PAO2 is entered in kPa)
Resolution: 0.1 or three significant digits (whichever is greater)
PaO
/
F
Ratio (P/F)
IO
2
2
The P
/ F
ratio is a simple assessment of gas exchange.
AO
IO
2
2
Range: 0 – 800 (P
Resolution: 0.1 or three significant digits (whichever is greater)

Waveforms and Loops

PCO
wave (capnogram)
2
Displays the CO
2
Maximum range: 0 – 150 mmHg (0 – 20 kPa)
PCO
/ Vte loop
2
Displays the patient's exhaled CO
phase, both values will be set to zero.
Maximum range (CO
Maximum range (Vte): 0 – 2.5 liters
Chapter 5: Volumetric Capnography
entered in mmHg) 0 – 106 (P
AO
2
value through the respiratory cycle as measured and reported by the CO
value on the vertical axis and exhaled Vt on the horizontal axis. During the inspiratory
2
): 0 – 150 mmHg (0 – 20 kPa)
2
entered in kPa)
AO
2
Operator's manual
sensor at the wye.
2
L2786 Rev. M