Insp Press/Insp Rise; Insp Time; Rate; Ncpap Flow - CareFusion AVEA Operating Instructions Manual

Hide thumbs Also See for AVEA:
Table of Contents

Advertisement

156
Insp Press
Range: 0 to 30 cmH
Resolution: 1 cmH
Default: 5 cmH
Accuracy: ±2 cmH
Insp Rise
Range: 1 to 9
(relative control with fast being a setting of 1, and slow a setting of 9)
Default
5
:
Note:
Inspiratory Rise is only available when the Rate is set.

Insp Time

0.15 to 3.0 seconds
Range
:
0.35 seconds
Default
:

Rate

Off, 1 to 80 bpm
Range
:
Off (nCPAP only)
Default
:

nCPAP Flow

Flow delivery is under software control and limited to a maximum of 15 LPM.
Advanced Settings
Note:
There are no advanced settings for the primary settings when the Rate is set to Off.
Alarms
Note:
The Alarms Settings window is not available in this mode when the Rate is set to Off. Existing machine alarms and
safety systems will be maintained. During nCPAP support with Rate set to Off, certain alarms will be suspended.
Chapter 6: Infant Non-invasive Ventilation
O
2
O
2
O
2
O
2
Operator's manual
L2786 Rev. M