AVEA ventilator systems
High nCPAP pressure
A high priority audible/visual alarm will be activated whenever the nCPAP exceeds the high airway pressure threshold for
a period greater than 15 seconds.
Threshold: Set CPAP Level + 3 cmH
Tolerance: ± 0.5 cmH
Note:
Threshold is automatically updated upon acceptance of a change in the nCPAP control setting.
Low nCPAP Pressure
A high priority audible/visual alarm will be activated whenever the nCPAP pressure falls below the low airway pressure
threshold for a period greater than 15 seconds.
Threshold:
Tolerance:
Note:
Threshold is automatically updated upon acceptance of a change in the nCPAP control setting.
nCPAP Pressure Limit
A high priority audible/visual alarm will be activated whenever the nCPAP pressure is greater than airway pressure limit
for 3 seconds. The alarm will deactivate when the nCPAP pressure drops below 4.5 cmH
Pressure Limit: 11 cmH
Set CPAP Level + Inspiratory Pressure + 3 cmH
Tolerance: + 0.5 cmH
L2786 Rev. M
Chapter 6: Infant Non-invasive Ventilation
O
2
O
2
Set nCPAP Level - 2 cmH
Set nCPAP Level - 1 cmH
± 0.5 cmH
O
2
O (nCPAP only, rate set to off)
2
O
2
O (if set nCPAP ≥ 3 cmH
2
O (if set nCPAP < 3 cmH
2
O (nCPAP / IMV, rate not zero)
2
O)
2
O)
2
O.
2