Hot Wire Flow Sensor Specifications - CareFusion AVEA Operating Instructions Manual

Hide thumbs Also See for AVEA:
Table of Contents

Advertisement

218

Hot Wire Flow Sensor Specifications

Table E-2 Hot wire sensor specifications
Appendix E: Sensor Specifications & Circuit Resistance
Part Number
Type:
Circuit Location:
Performance Specifications
Flow Range:
Vol. Accuracy:
Flow Resistance:
Dead Space:
Freq. Response*:
Calibration:
Linearity:
Operating Temperature:
Physical Specifications
Sensor length
Diameter Insp (Vent Side)
Diameter Exp (Patient Side)
Tube length
Connector
Weight
Service Life
Sterilization
Materials
51000-40081
Multiple use heated wire
Wye
0 (+/- 0.002) to 30 L/min
+/-10%
6 cmH
O @ 20 L/min
2
0.8 mL
16 Hz
36 point curve
< 2%
5 to 40
C
o
1.68"
15 mm OD
15 mm OD
N/A
Pin & Socket type
< 10g (not including wire)
25 cycles
Steam Autoclave
Sensor - Delrin
Wire – Platinum
Screen – Stainless Steel 304 or 316
Pin – PhBz, gold over nickel plated
Spacer - Delrin
Operator's manual
L2786 Rev. M