CareFusion AVEA Operating Instructions Manual page 149

Hide thumbs Also See for AVEA:
Table of Contents

Advertisement

AVEA ventilator systems
3.
Enable CO
Monitoring by touching the Enable/Disable button.
2
Figure 5–3: Monitoring Tab, Utility Screen
Note:
Capnography requires either a proximal flow sensor or circuit compliance compensation to be active.
If CO
monitoring is enabled but a proximal flow sensor or circuit compliance compensation is not active, an alert dialog
2
box appears.
Figure 5–4: Vco2 Alert Dialog
4.
If volumetric capnography is required, add a proximal flow sensor or enable circuit compliance compensation (or do
both), and then re-enable CO
monitor are available.
5.
Remove the appropriate airway adapter from its packaging and make sure it is undamaged and ready to use.
L2786 Rev. M
Chapter 5: Volumetric Capnography
monitoring as described above; otherwise, only the PCO
2
waveform and End-tidal CO
2
135
2