Appendix D: Monitor Ranges and Accuracies

DISPLAY
DESCRIPTION
VOLUME MONITORS
The volume measured during the inspiratory phase of the breath is accumulated as the inspired tidal volume, and the volume
measured during the exhalation phase is accumulated as the exhaled tidal volume. This volume does not include the volume
delivered by the Circuit Compliance Compensation function for volume breaths.
Vte
Exhaled tidal volume. Exhaled volume readings are
measured by the expiratory flow sensor. This
reading may be affected by the humidifier setting in
the ventilator. When using a proximal flow sensor,
the VTe is measured as the expiratory flow goes
away from the patient at the point of insertion of the
sensor (between the patient interface and the wye)
Vte/kg
Exhaled tidal volume adjusted for patient weight
Vti
Inspired tidal volume. VTi is measured by the
inspiratory flow sensor inside the ventilator and
reflects the volume without compensating for tubing
compliance. It is a calculation of the difference
between the flow delivered and the flow exhaled
during inspiration. When using a proximal flow
sensor, the VTi is measured as the inspiratory flow
(translated into volume) goes to the patient at the
point of insertion of the sensor (between the patient
interface and the wye).
Vti/kg
Inspired tidal volume adjusted for patient weight
Spon Vt
Spontaneous tidal volume. A zero (0) Spont Vt
indicates the most recent breath was not a
spontaneous breath. It is an instantaneous value.
Spon Vt/kg
Spontaneous tidal volume adjusted for patient
weight
Mand Vt
Mandatory tidal volume. Displayed as a rolling
average of either 8 breaths or one minute,
whichever occurs first.
Mand Vt/kg
Mandatory tidal volume adjusted for patient weight
Vdel
Total delivered machine volume measured by the
ventilator's inspiratory flow sensor. This value will
be greater than the VTi if tubing compliance
compensation is active.
% Leak
Percent leakage. The difference between the
inspired and exhaled tidal volumes in terms of %
difference.
Ve
Minute Volume. Volume of gas exhaled by the
patient during the last minute.
Ve/kg
Minute volume adjusted for patient weight
Spon Ve
Spontaneous minute volume.
Spon Ve/kg
Spontaneous minute volume adjusted for patient
weight
RATE/TIME MONITORS
Rate
Breath Rate.
Spon Rate
Spontaneous breath rate. Reflects spontaneous
rate for the last minute.
L2786 Rev. M
RANGE
ACCURACY
0 to 4 L
(± 20ml + 10% of reading)-Adult machine
sensor
(± 1 ml + 10% of reading)-Neonate wye
sensor
0 to 4 ml/kg
0 to 4 L
(± 20ml + 10% of reading)-Adult machine
sensor
(± 1 ml + 10% of reading)-Neonate wye
sensor
0 to 4 ml/kg
0 to 4 L
(± 20ml + 10% of reading)-Adult machine
sensor
(± 1 ml + 10% of reading)-Neonate wye
sensor
0 to 4 ml/kg
0 to 4 L
(± 20ml + 10% of reading)-Adult machine
sensor
(± 1 ml + 10% of reading)-Neonate wye
sensor
0 to 4 ml/kg
Derived
0 to 4L
(± 20ml + 10% of reading)-
Derived
Derived
0 to 99.9 L
Derived
0 to 999 ml/kg
Derived
0 to 99.9L
Derived
0 to 999ml/kg
Derived
0 to 200 bpm
± 3% or ± 2 bpm whichever is greater
0 to 200 bpm
± 3% or ± 2 bpm whichever is greater