Appendix E: Sensor Specifications & Circuit Resistance; Varflex ® Sensor Specifications - CareFusion AVEA Operating Instructions Manual

Hide thumbs Also See for AVEA:
Table of Contents

Advertisement

AVEA ventilator systems
Appendix E: Sensor Specifications & Circuit Resistance
VarFlex
Sensor Specifications
®
Table E.1 Varflex Flow Sensor Specifications
Sensor
Part Number
Type
Circuit Location
Performance Specifications
Flow Range
Diff Pres Range
Accuracy*
Resistance
Dead Space
Freq. Response**
Airway Pres Range
Calibration (EEPROM)
Linearity
Operating Temperature
Physical Specifications
Sensor Length
Diameter Insp (Vent Side)
Diameter Exp (Patient)
Tube Length
Connector
Weight
Service Life
Sterilization
Material
L/min: Dry air at 77° F (25° C) and 14.7 psig barometric pressure.
* Includes ± 1% for linearity and hysteresis with no zero drift for the pressure transducer and ± 2 % for temperature and
humidity variations.
The sensor must be corrected for barometric pressure, and oxygen concentration.
** Frequency Response is signal attenuation to 0.707 input and assumes 100 Hz sample rate.
L2786 Rev. M
Infant 15 mm
7002500
Single Use
Wye
0.024 to 30 L/min
± 5.72 cmH
O
2
± (0.012 L/min + 5% or reading
4.5 cmH
O @ 30 L/min
2
0.7 ml installed
17 Hz
-140 to 140 cmH
O
2
29 Point Curve
< 1% between points
5° to 40° C
41° to 104° F
1.36 in (3.5 cm)
15 mm OD
15 mm OD
48 in (121.9 cm)
Bicore Proprietary
22 g (0.7 oz)
Single Patient Use
NA
Sensor – Lexan
Flap – Mylar
Tubing – PVC
Connector - ABS
Adult 15 mm
7002300
Single Use
Wye
1.2 to 180 L/min
± 5.72 cmH
O
2
± (0.1 L/min + 5% or reading
2.4 cmH
O @ 60 L/min
2
9.6 ml installed
26 Hz
-140 to 140 cmH
O
2
29 Point Curve
< 1% between points
5° to 40° C
41° to 104° F
2.45 in (6.2 cm)
15 mm OD
15 mm OD
73 in (185.4 cm)
Bicore Proprietary
31 g (1.0 oz)
Single Patient Use
NA
Sensor – Lexan
Flap – Mylar
Tubing – PVC
Connector - ABS
217

Advertisement

Table of Contents
loading

Table of Contents