CareFusion AVEA Operating Instructions Manual page 236

Hide thumbs Also See for AVEA:
Table of Contents

Advertisement

222
AVEA MESSAGE BAR TEXT
"Printer Ready."
"Printer Busy."
"Volume Limit disabled."
"Proximal Flow Sensor disconnected."
"Flow sensor is not Heliox-compatible."
"Proximal Airway Line disconnected."
"Proximal Flow Sensor conflict.
"Esophageal monitoring not available."
"Tracheal monitoring not available."
"Flow Sensor Error."
"Wye Sensor Error."
"Device Error."
"Esophageal Balloon Leak Test Failed."
"Stopped: Patient Effort Detected"
"Proximal Flow Sensor Ready"
Appendix F: AVEA Message Bar Text
Sending screen data to printer has completed.
Print Screen button was pressed, device has not completed
sending data from previous activation.
On disconnect of WFS (Neo or Hotwire) when Size is NEO and
Volume Limit is active.
On disconnect of WFS, any type.
On connect of Hotwire WFS when Heliox is active.
On disconnect of Proximal Pressure connection.
On simultaneous connect of Hotwire and VarFlex WFS.
On connect of Esophageal Balloon when size is NEO.
On connect of Tracheal Catheter when size is NEO.
On power up, failure to validate any internal flow sensor.
On connect and failure to validate any proximal flow sensor.
On detection of a fault classified as "Device Error"
On failure of Esophageal Balloon leak test.
Upon detecting Patient effort in maneuvers which require a
passive patient
Operator's manual
CAUSE
L2786 Rev. M