Capnometry Alarms; High Etco2; Low Etco2 - CareFusion AVEA Operating Instructions Manual

Hide thumbs Also See for AVEA:
Table of Contents

Advertisement

AVEA ventilator systems
Alarms
Figure 5–7: Capnometry Alarms
High EtCO
2
Creates a low-priority alarm if the monitored EtCO
Range: 6 to 150 mmHg (0.8 – 20 kPa) or Off
Resolution: 1 mmHg (0.1 kPa)
Default: 60 mmHg (8 kPa)
Note:
The High EtCO
2
Low EtCO
2
Creates a low-priority alarm if the monitored EtCO
Range: 1 – 145 mmHg (0.1 – 19.3 kPa) or Off
Resolution: 1 mmHg (0.1 kPa)
Default: 30 mmHg (4 kPa)
Note:
The Low EtCO
alarm must be set at least 5 mmHg (0.7 kPa) below the High EtCO
2
L2786 Rev. M
Chapter 5: Volumetric Capnography
alarm must be set at least 5 mmHg (0.7 kPa) above the Low EtCO
exceeds this setting (see the previous figure).
2
does not exceed the setting (see the previous figure).
2
alarm setting.
2
alarm setting.
2
141