Checking the Accuracy of the CAPNOSTAT 5

The accuracy of the CAPNOSTAT 5 sensor should be compared against a calibration gas every twelve months.
Attach the end of the CO
1.
Figure 5–12: Bottom of UIM
2.
Attach the CO
sensor to the airway adapter.
2
3.
Access the Capnography Utilities by depressing the Screens button, selecting Utility, and the selecting the Monitoring
tab.
Figure 5–13: Screens Softkey
Chapter 5: Volumetric Capnography
sensor cable to the connection on the bottom of the AVEA UIM.
2
Screens softkey
Operator's manual
L2786 Rev. M