Volume Alarms

Low Exhaled Minute Volume (Low V
This is a high priority audible/visual alarm. LOW MINUTE VOLUME is displayed and a high priority tone sounds
whenever the monitored exhaled minute volume is less than the Low Exhaled Minute Volume threshold setting.
Range:
Defaults:
High Exhaled Minute Volume (High V
This is a medium priority audible/visual alarm. HIGH MINUTE VOLUME is displayed and a medium priority tone sounds
whenever the monitored exhaled minute volume is greater than the High Exhaled Minute Volume threshold setting.
Range:
Defaults:
Low Exhaled Tidal Volume (Low V
A high priority audible/visual alarm shall be activated, and LOW TIDAL VOLUME shall be indicated, whenever the
absolute monitored exhaled tidal volume does not exceed the Low Tidal Volume alarm threshold setting for the Low Vte
Sensitivity setting.
Range: Off (indicated by 0.00) to 3.00 L
Off (indicated by 0) to 1000 mL
Off (indicated by 0.0) to 300.0 mL (Neonate)
Resolution:
Accuracy:
Defaults:
Chapter 7: Alarms and Indicators
Off (indicated by 0), 1 to 50 L
Off (indicated by 0), 0.1 to 30.0 L
Off (indicated by 0), 0.01 to 5.00 L (Neonate)
1.00 Liter (Adults)
0.50 Liter (Pediatrics)
0.05 Liter (Neonate)
0 to 75 L
(Adult)
0.0 to 30.0 L
(Pediatric)
0.00 to 5.00 L
(Neonate)
30.0 L
(Adult/Pediatric)
5.00 L
(Neonate)
0.01 L (Adult)
1 mL
(Pediatric)
0.1 mL (Neonate)
± 0.01 L of monitored exhaled tidal volume (Adult)
± 1 mL of monitored exhaled tidal volume
± 0.1 mL of monitored exhaled tidal volume (Neonate)
0.00 L (Adult)
0 mL
(Pediatric)
0.0 mL (Neonate)
)
e
(Adult)
(Pediatric)
)
e
)
t
(Adult)
(Pediatric)
(Pediatric)
Operator's manual
L2786 Rev. M