Connecting The Test Fixture For Use; Connecting The Test Fixture With Apc-7 - Agilent Technologies 16191A Operation And Service Manual

Side electrode smd test fixture
Table of Contents

Advertisement

Connecting the Test Fixture for Use

Note
Some instruments require an adapter to be connected to the 16191A (see
\Applicable Instruments" in Chapter 1).
1. Calibrate your analyzer at the APC-7
Table 2-2 for the applicable error correcting function. After calibration, disconnect any
standards from the APC-7
2. Place the xture on the APC-7
3. Tighten (turn counterclockwise) the coupling nut of the APC-7
Figure 2-2. Connecting the Test Fixture with APC-7
connector plane before connecting test xture. See
R
connector.
R
connector.
R
connector.
R
Connector (4291B)
R
Preparation for Use
2-3