Saving/Recalling; Specifying A File; Saving And Recalling Instrument Status; Saving Measurement Data - Agilent Technologies E5052A Programmer's Manual

Signal source analyzer
Hide thumbs Also See for E5052A:
Table of Contents

Advertisement

Saving and Recalling (File Management)

Saving/Recalling

Saving/Recalling

Specifying a file

When running a command for saving, recalling, or managing files, use a file name with
extension to specify a particular file. Specify "A:" at the beginning of the file name when
specifying a file on the floppy disk. Also, when specifying a file name with a directory, use
"/" (slash) or "\" (backslash) as a delimiter.

Saving and recalling instrument status

You can save the instrument state by using one of the following methods:
Saving the entire instrument state (setting and data) into a file
Saving only the instrument setting into a file
Options for saving and recalling instrument status
When saving the instrument status into a file, you have two options for selecting the
content to be saved: entire instrument state (setting and data) and instrument setting only.
To select the content to be saved, use the following command:
:MMEMory:STORe:STYPe on page 288
To save the instrument setting (or instrument setting and data), use the following
command:
:MMEMory:STORe:STATe on page 287
Recalling a file saved with the above command reproduces the file's status when it was last
saved. To recall the settings from a file, use the following command:
:MMEMory:LOAD:STATe on page 282
Auto recall
The file saved with the name F:\autorec.sta or A:\autorec.sta will be automatically recalled
when the E5052A is powered on.

Saving measurement data

Measurement data (in a formatted data array) can be saved in the file in CSV (Comma
Separated Value) format. To save measurement data in a file, use the following command
("xx" specifies any one of the measurements (PN, SP, FP, TR, USER) while "y" specifies
the number of the trace):
:MMEMory:xx[1-1]:TRACe[1-y]:STORe
Executing the above command will save the specified trace data of the specified
measurement. Note that the data saved using the above command cannot be recalled from
the E5052A.
78